Two Limitations of Subcutaneous Implantable Cardioverter Defibrillator in the Same Patient Warranting Its Explant
Unusual clinical course, Unusual or unexpected effect of treatment
Rahul Dhawan1ADEF, Mansoor Ahmad1DE, Aravdeep Jhand1DE, Sumera Kanwal1DE, Adeel Jamil2DE, Faris Khan1ADEF*DOI: 10.12659/AJCR.928983
Am J Case Rep 2021; 22:e928983
Figure 1. (A) Subcutaneous implantable cardioverter defibrillator (S-ICD) interrogation showing inappropriate T-wave oversensing leading to appropriate shock by the device. (B) Chest X-ray prior to S-ICD removal showing newly implanted transvenous implantable cardioverter defibrillator (TV-ICD). (C) TV-ICD with appropriate sensing leading to appropriate therapy. (D) Electroanatomic map of left ventricle in left lateral view showing ablation points in basal to mid anterolateral wall.