Delayed Pacemaker Generator Pocket and Lead Primary Infection Due to Burkholderia Cepacia
Unusual or unexpected effect of treatment, Rare disease, Educational Purpose (only if useful for a systematic review or synthesis)
Fatma Ben Abid ABCDEF 1*, Hussam Al-Saoub E 1, Faraj Howadi E 1, Ahmed AlBishawi EF 2, Maliha Thapur E 1DOI: 10.12659/AJCR.904435
Am J Case Rep 2017; 18:855-858
Abstract
BACKGROUND: Recently, the use of cardiac implantable electrophysiological devices (CIEDs) has increased. Advances in medical technology, an increasingly aging population, increases in clinical indications, and expanded medical insurance coverage for these devices have all contributed to this trend. Infection is considered to be one of the most serious complications of CIEDs and carries a significant risk of morbidity and mortality. Although infection with Staphylococcus sp. accounts for the majority of cases, other bacteria have been implicated as causative agents of infection of CIEDs.
CASE REPORT: We report the first case of primary pacemaker generator pocket and lead infection due to Burkholderia cepacia (B. cepacia) in the state of Qatar. To our knowledge, there have been few cases of CIED infection due to B. cepacia previously reported in the literature.
CONCLUSIONS: This case raises awareness of B. cepacia as a potential opportunistic pathogen in CIED infection. The more rare bacteria require culture on special media to provide an early diagnosis to enable proper antimicrobial therapy to commence. Adherence to infection control standards during CIED insertion would reduce infection from B. cepacia.
Keywords: Burkholderia cepacia, Cardiac Resynchronization Therapy Devices, Defibrillators, Implantable, Heart Valve Prosthesis Implantation, Pacemaker, Artificial
Background
Opportunistic infection of cardiac implantable electrophysiological devices (CIEDs) CIED is a serious complication that is still on the rise despite the use of prophylactic antibiotics and the application of strict infection control measures [1,2]. Multiple risk factors, such as diabetes, hypertension, chronic renal failure, and coronary artery disease, have been reported to increase the risk of CIED infection [3].
The major causative organisms associated with CIED infection are
Case Report
A 70-year-old woman presented to the emergency department with a two-day history of swelling, redness, and pain at the site of her cardiac pacemaker pocket. She had history of diabetes and hypertension, and in 2003, she had suffered from a myocardial infarction. The patient had a subcutaneous left pre-pectoral implantation of a pacemaker 11 months prior to her current presentation, but the reason for this was unclear, as her previous medical records from another country were unobtainable.
Physical examination on admission showed swelling, erythema, warmth, and tenderness at the site of pacemaker pocket without dehiscence or discharge. Laboratory findings were as follows: white blood cell count (WBC) of 8.5×109/L (reference range 4.0–10.0×109/L); hemoglobin (Hb) 13 gm/dL (reference range 12.0–15.0 gm/dL); platelets 312×103/μL (reference range 150–400×103/μL); HbA1c 7.8% (reference range 4.8–6.0%); ESR 10 mm/h (reference range 0–29 mm/hr); C-reactive protein 5 mg/L (reference range, 0–5 mg/L). Tests of renal function, liver function, and blood coagulation profile were within normal range. Procalcitonin, a marker of sepsis, was 0.04 ng/ml (<0.5 ng/ml low risk of severe sepsis; >2.0 ng/ml high risk of severe sepsis). Three sets of blood cultures were negative. Trans-thoracic echocardiography was normal.
The diagnosis of superficial infection at the pocket site was considered, based on the clinical presentation and findings by the emergency room physician. The patient was started empirically on oral Augmentin, 1 gm twice daily, and discharged home. The patient was not referred to infectious diseases, cardiology or electrophysiology subspecialties for evaluation and management. Two weeks later, the patient returned with worsening swelling, pain, and redness without discharge or fever. However, at this time, the patient was seen by infectious disease, electrophysiology, and cardiology teams. Repeated routine blood workup and blood cultures were again negative. A left upper chest ultrasound was performed and showed a collection at the pacemaker pocket measuring 5.6×1.7×3.6 cm in size (Figure 1). A trans-esophageal echocardiography was performed. The pacemaker wire was visualized in position at the right atrium and right ventricle without any masses or vegetations attached to it. The patient was started empirically on vancomycin 20 mg/kg every 8 hours, cefepime 2 gm every 8 hours, and rifampicin 450 mg twice daily without improvement. Both the generator and wires were removed after two weeks of antibiotic treatment. Five days later, culture from the pocket pus and pacemaker lead grew
A diagnosis of pocket and lead pacemaker infection by
Discussion
Cardiac implantable electrophysiological devices (CIED) include all permanent pacemakers and implantable cardioverter-defibrillators (ICDs). The clinical indications for the use of CIEDs have increased over the last few decades with the advances in cardiac electrophysiology and increase in life expectancy [1]. CIED implantation may be associated with complications, of which, infection is considered to be a major complication that carries considerable morbidity, mortality, and significant financial burden to healthcare systems [2,3].
CIED of infection is defined as the presence of either local signs of inflammation at the generator pocket site, or the presence of lead or valvular vegetations, including the Duke criteria for infective endocarditis [3–5]. Our patient had pocket and lead infection without bacteremia. The incidence of pacemaker-related infection is greatest among cases of abdominal implantation, which can be as high as 19.9%, and lowest with pre-pectoral implantation, with infection rates of 0.13% [6]. Despite improved surgical techniques and the use of infection control measures as well as antibiotic prophylaxis, the frequency of CIED infections is increasing disproportionately to the rate of implantation [1,2]. For example, there was a 42% increase in the rate of CIED implantation between 1990 and 1999 accompanied by a 124% infection rate [1,2]. This finding was most likely attributed to the increase in the incidence of major comorbidities.
Several risk factors can play an important role in the development of CIED infection [3]. Patient factors, including the presence of diabetes mellitus, heart failure, and renal dysfunction have the strongest association with CIED infection [3,6,7]. Our patient had underlying diabetes mellitus, which could have been the main underlying risk factor for her CIED infection. In addition, peri-operative factors, such as the lack of antibiotic prophylaxis prior to device implantation, temporary pacing prior to permanent device placement, the presence of fever within 24 hours of implantation, operator inexperience, and prolonged operative time, can also increase the risk of infection [6]. Furthermore, device factors that can be associated with increased risk of infection, and include abdominal generator placement, the presence of more than two electrodes leads, epicardial lead placement, multiple previous device revisions, and a previous history of CIED infection, are risk factors for infection [6]. Device contamination could be primary, when the device and/or pocket itself is the source of infection during implantation, or secondary, due to bacteremia from a distant source [8]. A recent study showed that 25% of infections were early (less than 28 days), while 33% were delayed (29– 362 days) and 42% were late (more than 365 days); there was no difference in the type of infective organism [8]. Our patient had primary delayed pacemaker infection that was most likely to have been contaminated during the insertion.
The major causative organisms associated with CIED infection are
The management of CIED infection includes a combination of long-term administration of appropriate antibiotics and full device extraction [6]. Because
A patient reported in a previously published case of
Conclusions
Awareness is needed among primary healthcare physicians to avoid delay in diagnosis and mismanagement of CIED-associated infection, including rare opportunist infection. Bacterial cultures may require special media to reach an early diagnosis and to start appropriate antimicrobial therapy. This case report has been presented to raise awareness of
References:
1.. Greenspon AJ, Patel JD, Lau E, 16-year trends in the infection burden for pacemakers and implantable cardioverter-defibrillators in the United States: 1993 to 2008: J Am Coll Cardiol, 2011; 58(10); 1001-6, pmid: 21867833
2.. Voigt A, Shalaby A, Saba S, Rising rates of cardiac rhythm management device infections in the United States: 1996 through 2003: J Am Coll Cardiol, 2006; 48(3); 590-91, pmid: 16875991
3.. Sohail MR, Uslan DZ, Khan AH, Management and outcome of permanent pacemaker and implantable cardioverter-defibrillator infections: J Am Coll Cardiol, 2007; 49(18); 1851-59, pmid: 17481444
4.. Duan X, Ling F, Zhou L: J Hosp Infect, 2007; 67(4); 392-93, pmid: 18023925
5.. Rajasekharan SK, Ramesh SA: Pol J Microbiol, 2013; 62(3); 327-300, pmid: 24459841
6.. Sohail MR, Uslan DZ, Khan AH, Management and outcome of permanent pacemaker and implantable cardioverter-defibrillator infections: J Am Coll Cardiol, 2007; 49(18); 1851-59, pmid: 17481444
7.. Mela T, McGovern BA, Garan H, Long-term infection rates associated with the pectoral versus abdominal approach to cardioverter-defibrillator implants: J Am Coll Cardiol, 2001; 88(7); 750-53
8.. Said SA, Nijhuis R, Derks A, Droste H, Septic pulmonary embolism caused by infected pacemaker leads after replacement of a cardiac resynchronization therapy device: Am J Case Rep, 2016; 17; 507-11, pmid: 27435910
9.. Baddour LM, Epstein AE, Erickson CC, Update on cardiovascular implantable electronic device infections and their management a scientific statement from the American Heart Association: Circ J, 2010; 121(3); 458-77
10.. Camus C, Leport C, Raffi F, Sustained bacteremia in 26 patients with a permanent endocardial pacemaker: Assessment of wire removal: Clin Infect Dis, 1993; 17(1); 46-55, pmid: 8353245
11.. Chua JD, Wilkoff BL, Lee I, Diagnosis and management of infections involving implantable electrophysiologic cardiac devices: Ann Intern Med, 2000; 133(8); 604-8, pmid: 11033588
12.. Hussein AA, Baghdy Y, Wazni OM, Microbiology of cardiac implantable electronic device infections: JACC Clinical Electrophysiology, 2016; 2(4); 498-505
13.. Kodali A, Khalighi K, A case of late implantable cardiac device infection with aspergillus in an immunocompetent host: Am J Case Rep, 2015; 16; 520-53, pmid: 26250569
14.. Amin M, Gross J, Andrews C, Furman S: Pacing Clin Electrophysiol, 1991; 14(2); 152-54, pmid: 1706498
15.. Bhojraj S, Hamdulay Z, Ali M, Kumar P: Ind J Thorac Cardiovasc Surg, 2007; 23(1); 25-27
16.. Nannini EC, Ponessa A, Muratori R: Braz J Infect Dis, 2015; 19(5); 543-45, pmid: 26322722
17.. Antony B, Cherian EV, Boloor R, Shenoy KV: Indian J Path Microbiol, 2016; 59(2); 197-99, pmid: 27166040
In Press
Case report
Am J Case Rep In Press; DOI: 10.12659/AJCR.949976
Case report
Am J Case Rep In Press; DOI: 10.12659/AJCR.950290
Case report
Am J Case Rep In Press; DOI: 10.12659/AJCR.950607
Case report
Am J Case Rep In Press; DOI: 10.12659/AJCR.950985
Most Viewed Current Articles
07 Dec 2021 : Case report
17,691,734
DOI :10.12659/AJCR.934347
Am J Case Rep 2021; 22:e934347
06 Dec 2021 : Case report
164,491
DOI :10.12659/AJCR.934406
Am J Case Rep 2021; 22:e934406
21 Jun 2024 : Case report
113,090
DOI :10.12659/AJCR.944371
Am J Case Rep 2024; 25:e944371
07 Mar 2024 : Case report
59,175
DOI :10.12659/AJCR.943133
Am J Case Rep 2024; 25:e943133