Table 1 Description of defects in cases of instrument-related malfunction.
| Number of occurrences | Causes | |
|---|---|---|
| Preoperative | 3 cases | Cable damage (1 case)External damage (1 case)Unknown details (1 case) |
| Intraoperative | 15 cases | Cable damage (13 cases)Aperture failure (1 case)Other (1 case) |
| Postoperative | 29 cases | Cable damage (28 cases)Damage to cover (1 case) |