Medial Tibial Plateau Stress Fracture Following Navigated Total Knee Arthroplasty: Two Case Reports

Diagnostic / therapeutic accidents

Muthana M. Sartawi1ABCDEF, Hafizur Rahman2CDEF*, James M. Kohlmann1ABCDEF

DOI: 10.12659/AJCR.933005

Am J Case Rep 2021; 22:e933005

0 Comments

Abstract

BACKGROUND: Computer navigation in total knee arthroplasty has been gaining worldwide interest among orthopedic surgeons. While there is controversial data regarding its potential better clinical outcomes compared to conventional total knee arthroplasty, it has been shown to improve component and limb alignment reliability at a potential cost of increased complications. We present 2 case reports of medial tibial stress fracture through navigated tibial cutting block pinhole sites.

CASE REPORT: Both cases involved morbidly obese patients who underwent a navigated total knee arthroplasty. During surgery, there were no intraoperative concerns. Both knees were well aligned postoperatively and no unusual pain was reported. At 6 months after total knee arthroplasty, a periprosthetic fracture was initiated at a cutting block pinhole site with varus collapse of the tibial component. In both cases, the pinhole site was close to the medial tibial cortex and the primary tibial component collapsed into the varus, requiring revision to a stemmed component with allograft bone. For both patients, the revision arthroplasty continues to perform well. We believe cutting block design in combination with small tibias and elevated body mass index contributed to this complication.

CONCLUSIONS: Robotic-assisted total knee replacement has been shown to improve precision in component alignment. We caution against placing cutting block pinholes close to the medial tibial cortex, especially in morbidly obese patients with small tibias.

Keywords: Arthroplasty, Fractures, Stress, Knee, Arthroplasty, Replacement, Knee, Knee Joint, Knee Prosthesis, Obesity, Morbid, Osteoarthritis, Knee, Reproducibility of Results, Tibia

Background

Navigated total knee arthroplasty (TKA) was first introduced in the late 1990’s. Improvements in component alignment following navigated TKA as compared to conventional TKA have been reported [1–4] and navigation has been shown to reliably assess component and limb alignment intraoperatively during TKA [1,5,6]. This alignment reliability comes with the disadvantages of an increased operation time and possibly increased complication rate [4,7,8]. The incidence rate of periprosthetic fractures through tracking pin sites is 0.06% to 4.8% [1]. However, limited data exist in the current literature from case reports and case series on pin-related fractures [1].

We present 2 cases of medial tibial stress fracture through navigated tibial cutting block pinhole sites. Both patients had a primary TKA implanted with the aid of navigational component alignment and ligamentous balancing. As part of the routine operative technique, a medial 4-mm unicortical anchoring pin was used to secure the tibial jig in place. Both patients required revision knee arthroplasty. The treating surgeons had extensive experience with the navigation system and implants used for these 2 cases and were the operating surgeons for the primary and then revision surgery for each case.

Case Reports

CASE 1:

A 47-year-old morbidly obese woman (body mass index, BMI: 51.4 kg/m2) underwent a navigated cruciate retaining cemented left total knee replacement for varus osteoarthritis (Figure 1A). There was no immediate evidence of intraoperative complications. A size 3 femoral component, size 2 tibial base plate, a 27-mm inset patella, and an 11-mm polyethylene tibial insert were implanted. Postoperative coronal alignment was 0.5 degrees valgus (Figure 1B). Early postoperative recovery was uneventful. At follow-up 6 weeks postoperatively, she had significantly improved mobility and knee range of motion, but described medial pain. Review of her X-rays demonstrated a periprosthetic fracture initiated at a cutting block pinhole site with varus collapse of the tibial component (Figure 1C). She was revised to stemmed tibial and femoral components with totally stabilized constraint, structural and morcellized allograft bone to the proximal tibia, and a lateral retinacular release (Figure 1D). At 12 months, the revision arthroplasty continues to perform well.

CASE 2:

A 58-year-old obese woman (BMI: 31.8 kg/m2) underwent a navigated posteriorly stabilized reverse hybrid left total knee replacement for valgus inflammatory arthritis (Figure 2A). As part of an ongoing radiostereometric analysis study, she was randomized to receive an uncemented tibial component. There was no immediate evidence of intraoperative complications. A size 2 cemented femoral component, size 3 uncemented Peri-Apatite™ tibial base plate, a cemented 29-mm inset patella, and an 11-mm polyethylene tibial insert were implanted. Postoperative coronal alignment was 1 degree varus (Figure 2B). Early postoperative recovery was uneventful. At follow-up 1 year postoperatively, she reported having medial knee pain and had increasing varus deformity. Serial X-rays demonstrated gradual collapse of her tibial component into the varus, with angulation of the medial cortex at the site of the cutting block pinhole (Figure 2C). Infection was excluded by normal C-reactive protein (CRP), erythrocyte sedimentation rate (ESR), and knee aspirate. At 1 year after total knee replacement, she underwent revision surgery to a posteriorly stabilized stemmed tibial component and morcellized allograft to the proximal medial tibia (Figure 2D). At 3 months, the revision arthroplasty continues to perform well.

Discussion

Periprosthetic tibial fractures after TKA have been attributed to technical surgical details and to predisposing host or patient risk factors. Patient risk factors include osteoporosis, morbid obesity, female gender, advanced age, and neurologic disorders [9-11]. Comorbid conditions like chronic renal disease, corticosteroid use, and rheumatoid arthritis may result in abnormal bone remodeling, which can delay the resolution of iatrogenic cortical defects created during navigated pin placement [9]. Surgical details that have been attributed to postoperative periprosthetic fractures include improper component position, varus malalignment, and revision TKA considerations. Component loosening may also be a risk [12–15]. Navigation tracker pins have previously been implicated in tibial and femoral fractures following computer-assisted TKA [1,9,16–21]. Stress concentration near pinhole sites has been documented to increase fracture risk [22] and the pins used to secure the tibial cutting jig in unicompartmental knee arthroplasty have been reported to predispose to tibial plateau stress fracture [23–25]. Here, we report 2 cases of tibial plateau stress fracture as a complication of tibial cutting block pin placement in computer-navigated TKA.

The fractures were not noted intraoperatively or detected on initial postoperative X-rays. Both patients were allowed to bear weight as tolerated postoperatively with routine physiotherapy input. No unusual pain was reported. There was no history of stiffness or trauma. Both knees were well aligned postoperatively. Neither patient had been diagnosed with osteoporosis. Repetitive loading was considered to have resulted in stress fractures through the medial tibial cortex, which had been compromised by the placement of the tibial cutting block pins. Both patients were obese and increased BMI may have contributed. Interestingly, the first patient had sustained a fracture by 6 weeks postoperatively, while the second performed well for almost 1 year before reporting medial-sided knee pain. It is possible that the fixation of the tibial component (cemented in the first case and uncemented Peri-Apatite™-coated in the second case) played a role in this difference.

We believe the design of the navigation tibial cutting block in combination with smaller tibias is an important contributing factor. The pin sites for the navigation cutting block are more distal than the corresponding non-navigation block, placing the holes further down the medial flare of the tibia and therefore closer to its medial border. Furthermore, as tibial size decreases, the medial cutting block hole becomes closer to the medial cortex of the tibia. We feel this combination placed the medial pinhole too close to the medial tibial cortex. Biomechanical studies are required to further quantify this problem.

The cause of fracture complication identified in this report has directed us to select a more proximal medial hole on the navigation tibial cutting block when small tibial size is anticipated and to use a short-cemented stem to bypass the stress riser caused by a pinhole that is noticed at the time of surgery to be too medial (Figure 3). Placing the medial hole more proximally through the block to avoid medial pin placement, however, does not allow the cutting block to be easily repositioned more distally if the tibia needs to be recut. Navigation tibial cutting block design changes would make recutting the tibia easier and would have the potential to prevent users of the system from creating the medial pinhole stress risers.

Conclusions

Robotic-assisted total knee replacement has been shown to improve precision in component alignment. Due to the position of the medial pinhole sites in some of the robotic knee-cutting block designs, we recommend caution in utilizing those pin hole sites that may predispose patients to periprosthetic fractures, especially in morbidly obese patients with small tibias.

References:

1.. Smith TJ, Siddiqi A, Forte SA, Periprosthetic fractures through tracking pin sites following computer navigated and robotic total and unicompart-mental knee arthroplasty: A systematic review: JBJS Rev, 2021; 9(1); e20.00091

2.. Jones CW, Jerabek SA, Current role of computer navigation in total knee arthroplasty: J Arthroplasty, 2018; 33; 1989-93

3.. Laskin RS, Beksaç B, Computer-assisted navigation in TKA: Where we are and where we are going: Clin Orthop Relat Res, 2006; 452; 127-31

4.. Chauhan SK, Scott RG, Breidahl W, Beaver RJ, Computer-assisted knee arthroplasty versus a conventional jig-based technique. A randomised, prospective trial: J Bone Jt Surg – Ser B, 2004; 86; 372-77

5.. Stulberg SD, Loan P, Sarin V, Computer-assisted navigation in total knee replacement: Results of an initial experience in thirty-five patients: J Bone Jt Surg – Ser A, 2002; 84; 90-98

6.. Hankemeier S, Hufner T, Wang G, Navigated intraoperative analysis of lower limb alignment: Arch Orthop Trauma Surg, 2005; 125; 531-35

7.. Stulberg SD, Yaffe MA, Koo SS, Computer-assisted surgery versus manual total knee arthroplasty: a case-controlled study: J Bone Jt Surg, 2006; 88; 47-54

8.. Park SE, Lee CT, Comparison of robotic-assisted and conventional manual implantation of a primary total knee arthroplasty: J Arthroplasty, 2007; 22; 1054-59

9.. Massai F, Conteduca F, Vadalà A, Tibial stress fracture after computer-navigated total knee arthroplasty: J Orthop Traumatol, 2010; 11; 123-27

10.. Meek RMD, Norwood T, Smith R, The risk of peri-prosthetic fracture after primary and revision total hip and knee replacement: J Bone Jt Surg– Ser B, 2011; 93B; 96-101

11.. Canton G, Ratti C, Fattori R, Periprosthetic knee fractures. A review of epidemiology, risk factors, diagnosis, management and outcome: Acta Biomed, 2017; 88; 118-28

12.. Lotke PA, Ecker ML, Influence of positioning of prosthesis in total knee replacement: J Bone Jt Surg – Ser A, 1977; 59; 77-79

13.. Felix NA, Stuart MJ, Hanssen AD, Periprosthetic fractures of the tibia associ -ated with total knee arthroplasty: Clin Orthop Relat Res, 1997(345); 113-24

14.. Petje G, Landsiedl F, Stress fracture of the tibia after total knee arthroplasty: Arch Orthop Trauma Surg, 1997; 116; 514-15

15.. Rand JA, Coventry MB, Stress fractures after total knee arthroplasty: J Bone Jt Surg – Ser A, 1980; 62; 226-33

16.. Ossendorf C, Fuchs B, Koch P, Femoral stress fracture after computer navigated total knee arthroplasty: Knee, 2006; 13; 397-99

17.. Jung H-J, Jung Y-B, Song K-S, Fractures associated with computer-navigated total knee arthroplasty: J Bone Jt Surg, 2007; 89; 2280-84

18.. Bonutti P, Dethmers D, Stiehl JB, Case report: Femoral shaft fracture resulting from femoral tracker placement in navigated TKA: Clin Orthop Relat Res, 2008; 466; 1499-502

19.. Wysocki RW, Sheinkop MB, Virkus WW, Della Valle CJ, Femoral fracture through a previous pin site after computer-assisted total knee arthroplasty: J Arthroplasty, 2008; 23; 462-65

20.. Hoke D, Jafari SM, Orozco F, Ong A, Tibial shaft stress fractures resulting from placement of navigation tracker pins: J Arthroplasty, 2011; 26; 504.e5-e8

21.. Li CH, Chen TH, Su YP, Periprosthetic femoral supracondylar fracture after total knee arthroplasty with navigation system: J Arthroplasty, 2008; 23; 304-7

22.. Kim K, Kim YH, Park WM, Rhyu KH, Stress concentration near pin holes associated with fracture risk after computer navigated total knee arthroplasty: Comput Aided Surg, 2010; 15; 98-103

23.. Brumby SA, Carrington R, Zayontz S, Tibial plateau stress fracture: A complication of unicompartmental knee arthroplasty using 4 guide pinholes: J Arthroplasty, 2003; 18; 809-12

24.. Seon JK, Song EK, Yoon TR, Tibial plateau stress fracture after unicondylar knee arthroplasty using a navigation system: Two case reports: Knee Surg Sport Traumatol Arthrosc, 2007; 15; 67-70

25.. Yang KY, Yeo SJ, Lo NN, Stress fracture of the medial tibial plateau after minimally invasive unicompartmental knee arthroplasty: A report of 2 cases: J Arthroplasty, 2003; 18; 801-3

In Press

04 Mar 2024 : Case report  Belgium

Nivolumab-Induced Cytokine Release Syndrome: A Case Report and Literature Review

Am J Case Rep In Press; DOI: 10.12659/AJCR.941835  

05 Mar 2024 : Case report  USA

Intrapulmonary Shunting and Paradoxical Air Embolism in Liver Transplantation: A Case Report

Am J Case Rep In Press; DOI: 10.12659/AJCR.943042  

05 Mar 2024 : Case report  China (mainland)

Successful Treatment Strategy for Duplicate Inferior Vena Cava and Deep Venous Thrombosis: Filter Placement...

Am J Case Rep In Press; DOI: 10.12659/AJCR.942578  

05 Mar 2024 : Case report  China (mainland)

Transformation of Severe Aplastic Anemia into Donor Cell Leukemia after Allogeneic Hematopoietic Stem Cell ...

Am J Case Rep In Press; DOI: 10.12659/AJCR.943801  

Most Viewed Current Articles

07 Mar 2024 : Case report  USA

Neurocysticercosis Presenting as Migraine in the United States

DOI :10.12659/AJCR.943133

Am J Case Rep 2024; 25:e943133

0:00

10 Jan 2022 : Case report  Germany

A Report on the First 7 Sequential Patients Treated Within the C-Reactive Protein Apheresis in COVID (CACOV...

DOI :10.12659/AJCR.935263

Am J Case Rep 2022; 23:e935263

19 Jul 2022 : Case report  Saudi Arabia

Atlantoaxial Subluxation Secondary to SARS-CoV-2 Infection: A Rare Orthopedic Complication from COVID-19

DOI :10.12659/AJCR.936128

Am J Case Rep 2022; 23:e936128

23 Feb 2022 : Case report  USA

Penile Necrosis Associated with Local Intravenous Injection of Cocaine

DOI :10.12659/AJCR.935250

Am J Case Rep 2022; 23:e935250

Your Privacy

We use cookies to ensure the functionality of our website, to personalize content and advertising, to provide social media features, and to analyze our traffic. If you allow us to do so, we also inform our social media, advertising and analysis partners about your use of our website, You can decise for yourself which categories you you want to deny or allow. Please note that based on your settings not all functionalities of the site are available. View our privacy policy.

American Journal of Case Reports eISSN: 1941-5923
American Journal of Case Reports eISSN: 1941-5923