30 October 2026 : Case report  Mexico

Tibial Post Fracture in a Posterior-Stabilized Knee Prosthesis: A Case Report and Review of the Literature

Unusual clinical course, Challenging differential diagnosis, Diagnostic / therapeutic accidents

César Abraham Ávila Lozano ACEG 1, César Alejandro Jiménez-Aroche ORCID logo ABCDEFG 1*, Jacobo Kerbel Sutton ACEG 1, Iturbide Andrés Ponce de León Sandoval ACEG 1, José Eduardo Miranda-Mayén ORCID logo ACEG 1, Ady Haim Kahana-Rojkind ACEG 1, Juan José Gómez Amezcua ACEG 1

DOI: 10.12659/AJCR.954794

Am J Case Rep 2026; 27:e954794

Abstract

0:00

BACKGROUND: Tibial post fracture is an uncommon but likely underrecognized complication of posterior-stabilized (PS) total knee arthroplasty (TKA) that often remains undiagnosed because conventional radiographs may be unrevealing. We report a late tibial post fracture in a previously well-functioning PS TKA, highlighting the role of dynamic fluoroscopy in raising preoperative suspicion and the successful use of isolated polyethylene insert exchange in a carefully selected patient.

CASE REPORT: A 92-year-old man presented with acute mechanical pain, recurrent clicking, and subjective instability 16 years after primary PS TKA following a low-energy flexion-rotation movement. Plain radiographs showed no evidence of periprosthetic fracture, component loosening, or implant failure. Dynamic fluoroscopy during flexion, which reproduced the patient’s symptoms, demonstrated abnormal anterior translation of the femoral component relative to the tibial insert during knee flexion, raising suspicion of cam-post mechanism failure. Revision surgery confirmed a fracture of the polyethylene tibial post with chronic wear at its base; the metallic components remained well fixed, without gross malposition. Isolated polyethylene insert exchange with a thicker insert restored intraoperative stability. At 4.5 months postoperatively, the patient demonstrated complete resolution of mechanical symptoms, with full extension, 100° to 110° of flexion, and preserved varus-valgus and anteroposterior stability.

CONCLUSIONS: This case illustrates how dynamic fluoroscopy may raise preoperative suspicion of tibial post fracture in selected patients with symptomatic late PS TKA and inconclusive conventional radiographs. In our patient, isolated polyethylene insert exchange was effective after confirmation of preserved component fixation, alignment, and overall stability, supporting its use in appropriately selected cases.

Keywords: Case Management, Case Reports, Knee, Knee Prosthesis, Tibial Fractures

Add Comment 0 Comments

In Press

Case report  Greece

Giant Primary Intermuscular Hydatid Cyst in the Posterior Thigh Compartment: A Case Report and Review of th...

Am J Case Rep In Press; DOI: 10.12659/AJCR.954745  

Case report  Turkey

Subacute Ventricular Septal Rupture Following Elective Percutaneous Coronary Intervention for Chronic Coron...

Am J Case Rep In Press; DOI: 10.12659/AJCR.954370  

Case report  Poland

Carotid Web-Associated Ischemic Stroke in a Young Woman: Diagnostic Pitfalls and Endovascular Management

Am J Case Rep In Press; DOI: 10.12659/AJCR.953876  

Case report  South Korea

Real-Time Ultrasound-Guided Deep Acupuncture at Zhongwan With Gastric Antral Wall Localization: A Case Report

Am J Case Rep In Press; DOI: 10.12659/AJCR.955058  

Most Viewed Current Articles

07 Dec 2021 : Case report  USA 28,723,499

Edwardsiella tarda: A Classic Presentation of a Rare Fatal Infection, with Possible New Background Risk Fac...

DOI :10.12659/AJCR.934347

Am J Case Rep 2021; 22:e934347

06 Dec 2021 : Case report  Brazil 182,184

Lipedema Can Be Treated Non-Surgically: A Report of 5 Cases

DOI :10.12659/AJCR.934406

Am J Case Rep 2021; 22:e934406

21 Jun 2024 : Case report  China (mainland) 125,598

Intracranial Parasitic Fetus in a Living Infant: A Case Study with Surgical Intervention and Prognosis Anal...

DOI :10.12659/AJCR.944371

Am J Case Rep 2024; 25:e944371

0:00

08 May 2026 : Case report  USA 98,809

Cutaneous Allodynia Associated With GLP-1RA Tirzepatide for Weight Management: A Case Series

DOI :10.12659/AJCR.952158

Am J Case Rep 2026; 27:e952158

0:00

Your Privacy

We use cookies to ensure the functionality of our website, to personalize content and advertising, to provide social media features, and to analyze our traffic. If you allow us to do so, we also inform our social media, advertising and analysis partners about your use of our website, You can decise for yourself which categories you you want to deny or allow. Please note that based on your settings not all functionalities of the site are available. View our privacy policy.

American Journal of Case Reports eISSN: 1941-5923
American Journal of Case Reports eISSN: 1941-5923